Performance Features:
1. Ultralow energy resolution, better light element detection effect
It adopts large area Be window electric-cooling SDD detector with resolution as low as 139ev and all parameters better than national standard. With its good energy linearity, energy resolution, spectral features and higher peak-to-background ratio, SDD detector delivers better analytical precision on light elements like Si, S, Al.
2.Higher test accuracy, lower detection limit
Specialized sample chamber design. Rotary sample platform can test light elements more effectively. Vacuum sample chamber guarantees test pressure less than 10pa, thus enlarging test range and improve test accuracy.
3. Faster analysis, present test result within seconds
It adopts self-developed digital multichannel technology, with maximum linear count rate reaching 10kcps, to quicken analysis and improve stability.
4. One-click intelligent operation
Professional software integrating FP and EC algorithms facilitates test operation and shortens test time.
5. Automation
It is highly automated, with the functions of auto-opening, auto-switch of collimators and filters. Automatic spectrum stabilizing device is used to guarantee performance consistency.
6. Intensity correction method
Customers can set and add various test modes. The internal intensity correction method can correct the deviation caused by the different geometrical state and nonuniform density of test samples.
7. Perfect optical path system
The self-developed optical path system is featured by shorter optical path, less optical loss and better excitation effect.
8.Safety protection function
It has sound safety protection function;no X ray leakage when instrument shuts down
9.Safety Indication
The green light is on when the instrument is powered on; the yellow light will blink to indicate radiation when it is testing.
Technical Parameters:
Model: TR-TC3600K
Size: 669mm×540mm×874mm
Sanple chamber size: Φ40.5mm×39.5mmm
Weight: 90KG
Simultaneous analysis of element: more than 30 kinds of elements (more elements can be added based on customer’s demands)
Analytical Range:1 ppm-99.99% (it varies according to different materials)
Analytical Accuracy: 0.05%(for samples with content higher than 96%)
Measuring time: 10 – 100 seconds
Resolution: 140±5eV
Excitation source: 50KV/1000uA-Rh, W target X-ray tube and high voltage power supply
Ambient temperature: 15℃—30℃
Ambient humidity: ≤70%
Working voltage: AC 110V/220V
Detecting object:Standard pressed powder sample(size:40mm in diameter)
Technical Parameters
Model: TR-TC3600K
Size: 669mm×540mm×874mm
Sanple chamber size: Φ40.5mm×39.5mmm
Weight: 90KG
Simultaneous analysis of element: more than 30 kinds of elements (more elements can be added based on customer’s demands)
Analytical Range:1 ppm-99.99% (it varies according to different materials)
Analytical Accuracy: 0.05%(for samples with content higher than 96%)
Measuring time: 10 – 100 seconds
Resolution: 140±5eV
Excitation source: 50KV/1000uA-Rh, W target X-ray tube and high voltage power supply
Ambient temperature: 15℃—30℃
Ambient humidity: ≤70%
Working voltage: AC 110V/220V
Detecting object:Standard pressed powder sample(size:40mm in diameter)